11. Appendix A - Specifications

Modes

Conductivity
Temperature compensated conductivity
Salinity
Temperature

Conductivity

Range

Accuracy

Resolution

Frequency

0 - 49.99

µ

S*

±

0.5% full scale

0.01

µ

S

70 Hz

0 - 499.9

µ

S

±

0.5% full scale

0.1

µ

S

70 Hz

0 - 4999

µ

S**

±

0.5% full scale

1

µ

S

240 Hz

0 - 49.99 mS***

±

0.5% full scale

0.01 mS

1562 Hz

0 - 499.9 mS****

±

0.5% full scale

0.1 mS

1562 Hz

Salinity

Range

Accuracy

Resolution

0-80 ppt (NaCl)

±

2% or

±

0.1 ppt

0.1 ppt

Temperature

Range

Accuracy

Resolution

-5 - 95°C

±

0.1°C +1LSD

0.1°C

Temperature Compensation

Method

Linear

Ref. temp., °C

15 - 25

Temp. Coefficient

0 - 4%/°C

Cell constant, cm

-1

0.01
0.08 - 0.12
0.8 - 1.2
8 - 12

Display

LCD

Cell connector

7-pin Mini Din

Power

AC, 115V, 220V

Approvals

UL, CSA, CE

Environmental requirements

95% RH non-cond

Size

9 x 9.5 x 4.4 inches

22.9 x 24.1 x 11.2 cm

Weight

2.6 pounds

1.1 kg

* Requires a cell constant of K=0.01, K=0.1 or K=1.
** Requires a cell constant of K=0.1, K=1 or K=10.
*** Requires a cell constant of K=1 or K=10.
**** Requires a cell constant of K=10.