YSI 3100 User Manual

Page 12

Advertising
background image

8

To choose one of the measurement modes (temperature is always displayed), simply press and
release the [MODE] key. Carefully observe the small legends at the far right side of the LCD.

Conductuctivity

with °C

Temperature

Compensated

Conductivity

with °C

Salinity
with °C

If the instrument is reading Temperature compensated conductivity, the large numbers on the
display will be followed by either a

µ

S or mS and the small portion of the display will show the

o

C flashing on and off.

If the instrument is reading Conductivity, the large numbers on the display will be followed by
either a

µµµµ

S or mS, but the small portion of the display will show the

o

C NOT flashing.

10.62

mS

24.8

°C

NOTE: The instrument measures conductivity in uS/cm or mS/cm even though the /cm is

not shown on the display.

If the instrument is reading Salinity, the large numbers on the display will be followed by a ppt.

34.2

ppt

24.8

°C

NOTE: Temperature compensated conductivity and salinity modes cannot be used unless a

YSI 3200 series cell is connected. When using a YSI 3400 series cell (or equivalent)
with the 3232 cell adapter, these modes will display an error message (“LErr”) since
3400 series cells do not contain a temperature sensor.

°C Not flashing = Conductivity

°C Flashing = Temperature compensated

conductivity

Salinity

Advertising