Rerr – YSI 3100 User Manual

Page 18

Advertising
background image

14

NOTE: You may see an error message in some manual ranges if the range selected is not
adequate for the sample you are measuring.

rErr

mS

24.8

°C

If this happens, simply press and release the [ENTER] key again until a range is selected which
is suitable for your sample. If you get lost and don’t know if you’re in a manual range or
autoranging, simply turn the instrument off and back on. The instrument will default to
autoranging when first turned on.

The YSI Model 3100 has five possible ranges. The number of ranges available for use depends
on the current cell constant.

Cell

Constant

Range 1

0 - 49.99

µ

S/cm

Range 2

0 - 499.9

µ

S/cm

Range 3

0 - 4999

µ

S/cm

Range 4

0 - 49.99 mS/cm

Range 5

0 - 499.9 mS/cm

K=0.01

K=0.1

K=1

K=10

NOTE: Cells may be used beyond their normal range, but with instability and/or reduced
accuracy.

Range error message