YSI 3100 User Manual

Page 22

Advertising
background image

18

SYMPTOM

POSSIBLE CAUSE

ACTION

8.

Main Display reads “LErr”

In temperature compensated
conductivity mode, temperature
exceeds the values computed using
user defined temperature coefficient
and/or reference temperature.

In cell constant cal mode,
temperature exceeds the values
computed using user defined
temperature coefficient and/or
reference temperature.

The user has selected Temperature
Compensated Conductivity or
Salinity and the current cell does
not contain a temperature sensor.

Adjust user defined temperature
coefficient (see 4.2 Temperature
Coefficient
) or reference
temperature (see 4.3 Reference
Temperature
)

Use a YSI 3200 series cell or turn
off temperature compensation.

9.

Secondary Display reads “Err ra”

System has failed its RAM test
check procedure

Turn instrument OFF and back ON.

Return the system for service.

10.

Secondary Display reads “Err ro”

System has failed its ROM test
check procedure

Turn instrument OFF and back ON.

Return the system for service.

11.

Secondary Display reads “udr”

Current cell does not contain a
temperature sensor (such as YSI
3400 series).

Temperature is < -5

o

C

Use a YSI 3200 series cell if
temperature readings or
compensation are required

Read solution of higher temperature

Replace Cell/Cable assy

Return system for service

12.

Secondary Display reads “ovr”

Temperature is > 95

o

C

Read solution of lower temperature

Replace Cell/Cable assy.

Return system for service

13.

Secondary Display reads “rEr”

Temperature jumper is set to

o

F

and reading is >199.9

o

F but < 203

o

F

Set jumper to read

o

C.

Return system for service