SYMPTOM

POSSIBLE CAUSE

ACTION

8.

Main Display reads “LErr”

•

In temperature compensated
conductivity mode, temperature
exceeds the values computed using
user defined temperature coefficient
and/or reference temperature.

•

In cell constant cal mode,
temperature exceeds the values
computed using user defined
temperature coefficient and/or
reference temperature.

•

The user has selected Temperature
Compensated Conductivity or
Salinity and the current cell does
not contain a temperature sensor.

•

Adjust user defined temperature
coefficient (see 4.2 Temperature
Coefficient
) or reference
temperature (see 4.3 Reference
Temperature
)

•

Use a YSI 3200 series cell or turn
off temperature compensation.

9.

Secondary Display reads “Err ra”

•

System has failed its RAM test
check procedure

•

Turn instrument OFF and back ON.

•

Return the system for service.

10.

Secondary Display reads “Err ro”

•

System has failed its ROM test
check procedure

•

Turn instrument OFF and back ON.

•

Return the system for service.

11.

Secondary Display reads “udr”

•

Current cell does not contain a
temperature sensor (such as YSI
3400 series).

•

Temperature is < -5

o

C

•

Use a YSI 3200 series cell if
temperature readings or
compensation are required

•

Read solution of higher temperature

•

Replace Cell/Cable assy

•

Return system for service

12.

Secondary Display reads “ovr”

•

Temperature is > 95

o

C

•

Read solution of lower temperature

•

Replace Cell/Cable assy.

•

Return system for service

13.

Secondary Display reads “rEr”

•

Temperature jumper is set to

o

F

and reading is >199.9

o

F but < 203

o

F

•

Set jumper to read

o

C.

•

Return system for service