2.

Press both the [UP ARROW] and [MODE] keys at the same time. The large portion of the
display will show “PLA” flashing, indicating that platinization is in process.

PLA

3.

After the platinization process is complete (about 30 minutes), the 3100 will return to normal
mode. Remove the cell from the platinizing solution. If you want to stop the platinization
before 30 minutes have passed, press both the [UP ARROW] and [MODE] keys at the same
time to abort.

4.

Thoroughly rinse the cell with distilled or deionized water.

5.

Promptly return the platinizing solution to its container.