Temperature coefficient, Reference temperature – YSI 3100 User Manual

Page 16

Advertising
background image

12

4.2 Temperature

Coefficient

Follow these steps to modify the temperature coefficient of the Model 3100.

1.

Press and release the [DOWN ARROW] and [MODE] keys at the same time. The CAL
symbol will appear at the bottom left of the display and the large portion of the display will

show

1.91

%

(or a value set previously using Advanced Setup).

1.91

%

22.7

°C

CAL

2.

Use the [UP ARROW] or [DOWN ARROW] key to change the value to the desired new
temperature coefficient.

1.50

%

22.7

°C

CAL

3.

Press the [ENTER] key. The word “SAVE” will flash across the display for a second to
indicate that your change has been accepted.

4.

Press the [MODE] key two times to return to normal operation; the CAL symbol will
disappear from the display.

See Appendix B for charts of common salt solutions at various temperatures.

4.3 Reference

Temperature

Follow these steps to modify the reference temperature of the Model 3100.

1.

Press and release the [DOWN ARROW] and [MODE] keys at the same time.

1.91

%

22.7

°C

CAL

Temperature coefficient

CAL symbol

CAL symbol

New temperature coefficient