Measurement modes, 4 measurement modes, Save – YSI 3100 User Manual

Page 11

Advertising
background image

7

1.00

CAL

4.

Use the [UP ARROW] or [DOWN ARROW] keys to change the value to the desired new cell
constant.

0.10

CAL

5.

Press the [ENTER] key. The word “SAVE” will flash across the display for a second to
indicate that your change has been accepted. The 3100 will return to normal operation mode.

SAVE

3.4 Measurement

Modes

The Model 3100 is designed to provide four distinct measurements:

!

Conductivity -- A measurement of the conductive material in the liquid sample without regard

to temperature

!

Temperature Compensated Conductivity -- Automatically adjusts the reading to a calculated

value which would have been read if the sample had been at 25

o

C (or some

other reference temperature which you choose). See section 4. Advanced
Setup
. NOTE: Requires YSI 3200 series cell.

!

Salinity -- A calculation done by the instrument electronics, based upon the conductivity and

temperature readings. NOTE: Requires YSI 3200 series cell.

!

Temperature -- Always displayed.

NOTE: When you turn the Model 3100 off, it will “remember” which mode you used last and

will return to that mode the next time the instrument is turned on.

Cell constant

New cell constant