Troubleshooting, Error messages – YSI 3100 User Manual

Page 21

Advertising
background image

17

6. Troubleshooting

Error Messages

The instrument performs a Power On Self Test each time it is turned on. The following error
messages are provided to facilitate troubleshooting. They appear on the LCD when an error is
detected.

SYMPTOM

POSSIBLE CAUSE

ACTION

1.

Instrument will not turn on

Power supply

Instrument requires service

Check power supply and AC outlet

Return system for service

2.

Instrument will not calibrate

Incorrect calibration procedure

Cell needs cleaning

Instrument requires service

See 4.1 Cell Calibration

See 5. Maintenance

Return system for service

3.

Instrument readings are inaccurate

Calibration is required

Cell is contaminated

Temperature coefficient has been
set incorrectly

Reference temperature incorrect

Readings are or are not temperature
compensated.

See 4.1 Cell Calibration

See 5. Maintenance

See 4.2 Temperature Coefficient

See 4.3 Reference Temperature

See 3.4 Measurement Modes

4.

Main Display reads “OVEr”

Conductivity Reading is over
range:

>112 uS with K=0.01 cell

>11.2 mS with K=0.1 cell

>112 mS with K=1 cell

>499.9 mS with K=10 cell

Salinity reading is > 80ppt

User cell constant cal is over the
limit of the current range

In all cases, check calibration
values and procedure; check
Advanced Setup settings.

Set cell constant to correct range.
See 3.3 Configure the 3200, Cell
Constant
.

If each of these is set correctly,
return system for service.

5.

Main Display reads “Undr”

User cell constant cal is under the
limit of the current range

Set cell constant to correct range.
See 3.3 Configure the 3200, Cell
Constant
.

Recalibrate using known good
conductivity standard. See 4.1 Cell
Calibration.

Follow cell cleaning procedure. See
5. Maintenance.

6.

Main Display reads “rErr”

User has selected manual ranging &
sample exceeds selected range

Conductivity reading is over the
range of the instrument: >499.9 mS

Use the ENTER key to select a
higher or lower manual range, or to
set system to Autoranging.

7.

Main Display reads “PErr”

Incorrect sequence of key strokes

Refer to manual section which
provides step by step procedures
for the function you are attempting.