MIQ/Blue PS

3 - 4

ba76095e02

01/2013

Special features with

certain configurations

 If, except in the MIQ/Blue PS, no other terminator switch is available

on the IQ S

ENSOR

N

ET

island, it must be set to ON. In addition, a ter-

minating resistor (included in the scope of delivery) must be con-
nected to one of the three SENSORNET connectors. For more
details, see page 3-6.

 If a DIQ/S 182 Universal Transmitter is on the IQ S

ENSOR

N

ET

island,

all terminator switches on the IQ S

ENSOR

N

ET

island must be set to

ON.

Example of determining

the main line

The following figure shows the main line L made up of the partial
lengths L1, L2 and L3 as the cable section L4 is shorter than L3:

Length of the main line = L1 + L2 + L3

Fig. 3-2

Example of determining the length of the main line

L1

= 150 m

L4 = 200 m

L2 = 200 m

L3 = 300 m