Installation
MIQ/Blue PS
3 - 6
ba76095e02
01/2013
Additional terminating
resistor on the MIQ/
Blue PS
An additional terminating resistor is only required if, except in the MIQ/
Blue PS, no other terminator switch is available on the IQ S
ENSOR
N
ET
island. The terminating resistor is connected to one of the three SEN-
SORNET connectors (Fig. 3-4). The terminating resistor must bridge
the two outside terminals of the SENSORNET connector. The SEN-
SORNET connector can still be used to connect a sensor.
Fig. 3-4
MIQ/Blue PS with additional terminating resistor
6
8
7
9
2
1
3
5
4
SN terminator
switch ON
additional terminating resistor