Warning – YSI MIQ/Blue PS Set User Manual
Page 29
Advertising
MIQ/Blue PS
Installation
3 - 19
ba76095e02
01/2013
Fig. 3-13 Line power connection.
Note
The complete assignment of the terminal strip is shown in section 3.7.
Warning
No free wires must be allowed to project into the enclosure. Oth-
erwise, there is a danger that areas safe to contact could come
into contact with dangerous voltages. Always cut off any wires
that are not in use as closely as possible to the cable gland.
9
Connect phases L and N to the terminal strip. Make sure that
the cable assignment agrees with the specification on the ter-
minal label under the terminal strip.
10
Tighten the coupling ring (pos. 3).
11
Close the enclosure.
3
N
L
Terminal
labeling:
X13
X12
100...
240V AC
NETZ/MAINS
L1
N