3 commissioning, 1 scope of delivery, 2 installation – YSI IQ S ENSOR N ET TriOxmatic 701 IQ User Manual

Page 11: Commissioning -1, Scope of delivery -1, Installation -1, 3commissioning

Advertising
background image

TriOxmatic

®

701 IQ

Commissioning

3 - 1

ba76047e01

01/2012

3

Commissioning

3.1

Scope of delivery

 TriOxmatic

®

701 IQ

 ZBK 25i accessory case

 The sensor is filled with electrolyte and provided with a protective

hood

and protective caps.

 Operating manual.

3.2

Installation

Connection cable

The SACIQ sensor connection cable is required to connect the sensor.
Information on this and other IQ S

ENSOR

N

ET

accessories is given in

the YSI catalog and on the Internet.

Note
How to connect the SACIQ sensor connection cable to the terminal
strip of an MIQ module is described in chapter 3 I

NSTALLATION

of the

IQ S

ENSOR

N

ET

system operating manual.

Are the plug

connections dry?

Before connecting the sensor and sensor connection cable, please
make sure that the plug connections are dry. If moisture gets into the
plug connections, first dry the plug connections (dab them dry or blow
them dry using compressed air).

Note
Do not suspend the sensor on the sensor connection cable. Use a
sensor holder or an armature. Information on this and other
IQ S

ENSOR

N

ET

accessories is given in the YSI catalog and on the

Internet.

Minimum approach flow

The minimum required approach flow at the sensor must be present
during measurement (see chapter 7 T

ECHNICAL

DATA

). The minimum

approach flow can, e.g. be achieved by:

 the flow rate of the water

 turbulence in the activated sludge tanks in the vicinity of the sensor.