Maintenance, cleaning, disposal, and replacement
TriOxmatic
®
701 IQ
5 - 18
ba76047e01
01/2012
3
Clamp the sensor in a stand.
Submerse the electrode unit in the RL-AG/Oxi cleaning
solution to just above the silver counter electrode.
4
Leave the cleaning solution to work for 1 hour.
5
Rinse the sensor head and electrode unit several times with
deionized water.
6
Rinse the sensor head and electrode unit for at least an hour in
deionized water.