YSI IQ S ENSOR N ET TriOxmatic 701 IQ User Manual

Page 30

Advertising
background image

Maintenance, cleaning, disposal, and replacement

TriOxmatic

®

701 IQ

5 - 8

ba76047e01

01/2012

Caution

The ELY/A electrolyte solution irritates eyes, skin and mucus
membranes. If it comes into contact with the eyes, rinse
thoroughly with water and consult a doctor! During working
activities, always wear suitable protective gloves and protective
goggles/face shield! Follow the safety datasheet.

10

Remove any air bubbles by carefully tapping the membrane
head.

11

Screw the membrane head onto the shaft while holding the
sensor at an angle. Excess electrolyte solution is forced out of
the ventilation area.

12

The filling should be free of air bubbles as far as possible.
However, small air bubbles do not cause any interference.

13

After approx. 60 minutes polarization time, the sensor is ready
for operation.

14

Recalibrate the sensor (see section 4.2 C

ALIBRATION

).

Ventilation area (see above)