YSI IQ S ENSOR N ET TriOxmatic 701 IQ User Manual

Page 22

Advertising
background image

Measuring / Operation

TriOxmatic

®

701 IQ

4 - 6

ba76047e01

01/2012

Measured values within

the required precision

If the measured value lies within the range of precision required by the
user, no cleaning or recalibration is necessary.

Example:
Required precision 5 %.
Measured value in air: 97 % saturation (nominal value: 100%)

 No cleaning or recalibration is necessary as the measurement error
lies within the range of precision required by the user.

Function check in air-

saturated water

In the case of air temperatures under 5 °C do not perform the function
check in air but in air-saturated water that has a higher temperature.
You obtain air-saturated water by pouring water several times in and
out of two vessels so that it sparkles.
Then submerse the sensor in a vessel with air-saturated water and
perform the function check as described above.

Advertising