Maintenance, cleaning, disposal, and replacement
TriOxmatic
®
701 IQ
5 - 16
ba76047e01
01/2012
10
Rinse the sensor head and electrode unit several times with
deionized water.
11
Rinse the sensor head and electrode unit for at least an hour in
deionized water.
12
Carefully shake off the drops of water.
13
Fill a new WP 25i membrane cap and screw it on (see section
5.3 C
).
14
Connect the sensor again with the SACIQ sensor connection
cable.