Maintenance, cleaning, disposal, and replacement
5 - 19
ba76047e01
01/2012
7
Carefully shake off the drops of water.
8
Fill a new WP 25i membrane cap and screw it on (see section
5.3 C
).
9
Connect the sensor again with the SACIQ sensor connection
cable.
10
After approx. 60 minutes polarization time, the sensor is ready
for operation.
11
Recalibrate the sensor (see section 4.2 C
).