TriOxmatic

®

701 IQ

Maintenance, cleaning, disposal, and replacement

5 - 17

ba76047e01

01/2012

Cleaning the silver

counter electrode in a

beaker

Caution

The reference electrode must not come into contact with the
cleaning solution under any circumstances. This could destroy
the reference electrode and cause the sensor to become
defective.

15

After approx. 60 minutes, the sensor is ready for operation.

16

Recalibrate the sensor (see section 4.2 C

ALIBRATION

).

1

Wipe the silver counter electrode with a lint-free paper towel
and carefully remove any loose deposits.

2

Fill a beaker (150 ml, high-sided form) with approx. 25 ml RL-
AG/Oxi cleaning solution.