What to do if...

TriOxmatic

®

701 IQ

6 - 2

ba76047e01

01/2012

Measured values too

high

Incorrect temperature

display

Cause

Remedy

– Sensor not yet completely

polarized

– Wait for complete

polarization (at least 1 hour)

– Sensor has not been calibrated

for a long time

– Clean the outside of the

sensor, then recalibrate
(section 5.2 and section 4.2)

Cause

Remedy

– Temperature sensor defective

– Return the sensor