YSI IQ S ENSOR N ET TriOxmatic 701 IQ User Manual

Page 13

Advertising
background image

TriOxmatic

®

701 IQ

Commissioning

3 - 3

ba76047e01

01/2012

3.3

Commissioning / Getting the instrument ready for
measuring

Note
For the sensor to polarize, the following conditions must be fulfilled:

 The sensor is connected with the IQ S

ENSOR

N

ET

system via the

SACIQ sensor connection cable.

 The IQ S

ENSOR

N

ET

system is in operation

 The sensor has been identified by the IQ S

ENSOR

N

ET

system.

Precision

measurements

Recommendation: In order to carry out precision measurements, leave
the sensor to polarize for a longer period of time, e.g. overnight, and
recalibrate it the following day.

1

Pull the protective cap off the protective hood of the sensor.

2

Leave the sensor to lie in the air for at least 60 minutes while it
is switched on (polarization).

3

If required, assign a user-defined name to the sensor (see
relevant IQ S

ENSOR

N

ET

system operating manual).

4

Set the sensor (see section 3.4).

5

Calibrate the sensor (see section 4.2 C

ALIBRATION

).