3 calibration history, 4 reactivation of the last valid calibration, Calibration history -4 – YSI IQ S ENSOR N ET TriOxmatic 701 IQ User Manual

Page 20: Reactivation of the last valid calibration -4

Advertising
background image

Measuring / Operation

TriOxmatic

®

701 IQ

4 - 4

ba76047e01

01/2012

4.2.3

Calibration history

Calibration history

(available in the

IQ S

ENSOR

N

ET

system 2020 XT)

Fig. 4-1

Calibration history TriOxmatic

®

701 IQ

The calibration history contains the following information:

 Date of the calibration

 Relative Slope (dimensionless)

 Rating of the calibration:

o.k.: Calibration

was

successful.

The new calibration values are used for the measurements.

Error: Calibration was not successful.

The Sensor is blocked for measurement.

4.2.4

Reactivation of the last valid calibration

The TriOxmatic

®

701 IQ enables you to reactivate the last valid

calibration when needed. Thereby, it is possible to continue with
measurements, when a calibration failed or it is assumed that optimal
calibration conditions were not met.

Note
The reactivation of former calibration data is only a temporary solution.
Please consider, that thereby possibly wrong measurement values are
produced. Please check correct functioning of the sensor by performing
a function check and/ or a user calibration.

Reactivating the last

valid calibration data

actual active
calibration

chronological list
with the last
calibrations

1

Open the the setting table (see section 3.4).

2

Choose in the menue item Calibration the setting last valid and
quit the setting table afterwards with Save and quit.