3 changing the electrolyte and membrane cap, Changing the electrolyte and membrane cap -5, Ead (section 5.3 – YSI IQ S ENSOR N ET TriOxmatic 701 IQ User Manual

Page 27

Advertising
background image

TriOxmatic

®

701 IQ

Maintenance, cleaning, disposal, and replacement

5 - 5

ba76047e01

01/2012

5.3

Changing the electrolyte and membrane cap

YSI delivers the sensor ready for operation. The electrolyte solution
and the membrane head must only be replaced:

 the membrane is heavily contaminated and a calibration error

occurs
(log book message)

 the membrane is damaged

(log book message)

 the electrolyte solution is depleted or the counter electrode is

poisoned
(log book message)

 after the gold working electrode and the silver counter electrode

have been cleaned

Note
Information on the contents and structure of the log book and how you
can call it up is given in the L

OG

BOOK

chapter of the IQ S

ENSOR

N

ET

system operating manual.

Note
Only unscrew the membrane head for maintenance purposes. After
doing so, always use a new membrane head!

Changing the electrolyte

and membrane cap

Caution

The ELY/A electrolyte solution irritates eyes, skin and mucus
membranes. If it comes into contact with the eyes, rinse
thoroughly with water and consult a doctor! During working
activities, always wear suitable protective gloves and protective
goggles/face shield! Follow the safety datasheet.

1

Pull the sensor out of the sample.

2

Get rid of any coarse contamination on the sensor (e.g. brush
it off in a bucket of tap water, wash it down with a hose or wash
it off with a cloth).

3

Unscrew the protective hood from the sensor.

Advertising