4 electrical data, 5 characteristic data on delivery, Electrical data -4 – YSI IQ S ENSOR N ET TriOxmatic 701 IQ User Manual

Page 50: Characteristic data on delivery -4

Advertising
background image

Technical data

TriOxmatic

®

701 IQ

7 - 4

ba76047e01

01/2012

7.4

Electrical data

7.5

Characteristic data on delivery

Zero signal

< 0.05 % of the saturation value

Response time at 25 °C

t

90

(90 % of the final value display after) < 30 s

t

99

(99 % of the final value display after) < 90 s

Own consumption

0.027 µg h

-1

(mg/l)

-1

at 20 °C

Drift

< 3 % per month during permanent polarization

Working life per

electrolyte filling

Nominal voltage

Max. 24VDC
via the IQ S

ENSOR

N

ET

(for more

details, see chapter T

ECHNICAL

DATA

of the IQ S

ENSOR

N

ET

system operating manual)

Power consumption

0.2 W

Protective class

III

Oxygen concentration

Theoretical electrolyte reserve

Air saturation

approx. 1.5 years

< 2 mg/l

approx. 7.5 years