2 info messages, Info messages -2 – YSI IQ S ENSOR N ET TriOxmatic 701 IQ User Manual

Page 54

Advertising
background image

Indexes

TriOxmatic

®

701 IQ

9 - 2

ba76047e01

01/2012

9.1.2

Info messages

EI1332

Operational voltage too low
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components, replace components

EI2332

Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components,
replace components

ES1332

Component hardware defective
* Contact service

ESB332

SensReg: Electrolyte supply is depleted
* Change electrolyte solution and membrane head immediately (see
operating manual)

ESC332

SensLeck: Membrane head damaged
* Change electrolyte solution and membrane head immediately (see
operating manual)

Message code

Message text

Message code

Message text

IC1332

Sensor has been successfully calibrated
* For calibration data, see calibration history

IC4332

Last valid calibration has been activated. Make sure the sensor
operates correctly.

II1332

Language not available,
Default language German
* Contact service