4 tetracon® 700 iq (sw) setting table, Tetracon – YSI IQ SensorNet TetraCon Sensor 700 IQ (SW) User Manual

Page 13

Advertising
background image

TetraCon

®

700 IQ (SW)

Commissioning

3 - 3

ba76018e01

01/2012

3.4

TetraCon

®

700 IQ (SW) setting table

Menu item

Selection/values

Explanations

Measuring mode

Conductivity /cm

Salinity

TDS

Conductivity /m

Measured parameter in the measured
value display
(TDS = total dissolved solids)

Measuring ranges
with
Measuring mode,
Conductivity /cm

AutoRange

0 .. 20.00

m

S/cm

0 .. 200.0

m

S/cm

0 .. 2000

m

S/cm

0 .. 20.00 mS/cm

0 .. 200.0 mS/cm

0 .. 500.0 mS/cm

These measuring ranges are available for
selection. If the AutoRange menu item is
selected, the measurement range selection
and switchover is made automatically.

Measuring ranges
with
Measuring mode,
Conductivity /m

AutoRange

0 .. 2.000 mS/m

0 .. 20.00 mS/m

0 .. 200.0 mS/m

0 .. 2000 mS/m

0 .. 20.00 S/m

0 .. 50.00 S/m

Measuring range in
Measuring mode
Salinity

0 .. 70

The measuring range is permanently set.

Measuring range in
Measuring mode
TDS

0 .. 2000 mg/l

The measuring range is permanently set.

Temperature mode

°C

°F

Unit of the measured temperature value
(Celsius, Fahrenheit).

Temp. compensation

none

nonlinear

linear with setting

For natural waters (groundwater,
surface water, drinking water),
salinity (seawater) according to IOT

Other aqueous measuring solutions