YSI IQ SensorNet TetraCon Sensor 700 IQ (SW) User Manual

Page 14

Advertising
background image

Commissioning

TetraCon

®

700 IQ (SW)

3 - 4

ba76018e01

01/2012

Carrying out settings

Using

s

, switch from the measured value display to the main menu of

the settings. Then navigate to the setting menu (setting table) of the
sensor. The exact procedure is given in the relevant IQ S

ENSOR

N

ET

system operating manual.

Temp. compensation
linear

0.5 ... 3.0 %/K

Factor for linear temperature
compensation. This menu item only
appears if linear temperature
compensation is selected.

Reference
temperature

Tref20

(20 °C)

Tref25

(25 °C)

The reference temperature is the basis for
calculating the temperature compensation.

TDS factor

0,40 .. 1,00

Factor for automatic calculation of the total
dissolved solids by the sensor.

Cell constant

0.826 ... 1.008 cm

-1

Here you can set the cell constant if this is
necessary for a special application, e.g.
when using a flow-thru vessel.

Temperature
adjustment

-1.5 ... +1.5 K

Here you can balance the temperature
sensor in the sensor against a reference
temperature measurement.

Save and quit

The system confirms the saving of the
settings and the display switches to the
next higher level.

Quit

The display switches to the next higher
level without saving the new settings.

Advertising