6 what to do if, What to do if -1, 6what to do if – YSI IQ SensorNet TetraCon Sensor 700 IQ (SW) User Manual

Page 21

Advertising
background image

TetraCon

®

700 IQ (SW)

What to do if...

6 - 1

ba76018e01

01/2012

6

What to do if...

No temperature display

and/or no conductivity

display

Measurement does not

function

Measurement provides

implausible measured

values

Cause

Remedy

– System setting incorrect

– Correct the system setting

– Temperature sensor or

conductivity sensor defective

– Return the conductivity sensor

Cause

Remedy

– Protective cap still on

conductivity sensor

– Pull off protective cap

– System setting incorrect

– Correct the system setting

Cause

Remedy

– Conductivity sensor heavily

contaminated

– Clean conductivity sensor

– Boundary field not

maintained

– The conductivity sensor must

be surrounded by a gap of at
least 5 cm at the base and
sides when measuring at the
electrodes. Otherwise, the cell
constant can change (see
section 4.2.2)

– Electrodes damaged

– Return the sensor

– System setting incorrect

– Correct the system setting

– Measuring range exceeded

– Make sure the correct sensor is

being used for the application

– The sensor was installed in

an armature and the
boundary field is not sufficient

– Set the cell constant to the

value of the installed state (if
known)

– If the cell constant of the sensor

in the installed state is not
known, set the measured value
to the nominal value of a
sample (see section
4.2 A