Maintenance, cleaning, disposal
TetraCon
®
700 IQ (SW)
5 - 2
ba76018e01
01/2012
to protect the electrical contacts from corrosion. The protective cap
is available as an accessory under the order number 480 065. It is
included in the standard scope of delivery of the SACIQ SW sensor
connection cable.
Cleaning
5.3
Disposal
We recommend to dispose of the sensor as electronic refuse.
Contamination
Cleaning agents
Reaction time at
room temperature
Water-soluble
substances
Tap water
Any
Fats and oils
– Warm water and
household
detergent;
– In the case of
heavy
contamination:
Methylated spirits
– Any
– Maximum of 5
minutes
Lime and hydroxide
deposits
Acetic acid (10 %)
max. 5 minutes