Section 4.2.2 – YSI IQ SensorNet TetraCon Sensor 700 IQ (SW) User Manual

Page 17

Advertising
background image

TetraCon

®

700 IQ (SW)

Measuring / Operation

4 - 3

ba76018e01

01/2012

4.2.2

Determining the cell constant in a user-specific measuring
environment

1

Immerse the operable conductivity sensor in the test sample in
the measuring environment and wait until the measured value
is stable.

2

Read the conductivity on the display and note it down (->

χ

D

).

3

Take a representative sample simultaneously with the
conductivity measurement and from the immediate vicinity of
the sensor if possible.

4

Determine the conductivity of the sample without the
influence of boundary fields
(->

χ

X

). The measurement can,

for example, be performed as follows:

Measuring in the laboratory using a laboratory conductivity
measuring cell

Measuring using the TetraCon

®

700 IQ (SW) while

observing the bottom and lateral gaps according to section
4.1 M

EASURING

.

Attention: Set the same procedure for the temperature
compensation as for measuring the

χ

D

!

5

Read and note down the currently set cell constant in the
setting menu of the sensor (see section 3.4) (-> K

A

).

6

Calculate the new cell constant K

N

according to:

7

Set and store the new cell constant K

N

(see section 3.4).

K

N

=(

χ

X

/

χ

D

)·K

A