4.2.2

Calibration with CAL TEC AUTO

Messages displayed during calibration with CAL TEC AUTO

Display

Explanation

* Have any two technical buffer
solutions ready.

You can use any two different
buffer solutions to do this.
Confirm with

g

.

* Rinse the sensor.
* Immerse the sensor in the first
buffer solution.
* Wait for a stable measured
value.

Follow the instructions on the
display. As soon as a stable
measured value is reached, the
next display appears.

* Rinse the sensor.
* Immerse the sensor in the
second buffer solution.
* Wait for a stable measured
value.

Follow the instructions on the
display. As soon as a stable
measured value is reached, the
next display appears.

Successfully calibrated.
End of the CAL TEC AUTO
calibration.

The values determined for Slope
and Asymmetry potential are
displayed.
The calibration is complete.
Confirm with

g

.

The display returns to the
measured value display.