Measuring / Operation

SensoLyt

®

700 IQ (SW)

4 - 4

ba76006e01

01/2012

4.2.3

Calibration with CAL CON 2P

Messages displayed during two-point calibration with CAL CON 2P

Display

Explanation

* Have buffer pH 7.0 ± 0.5 and
any second buffer solution ready.

You can use two buffer solutions
of which the pH values according
to the actual temperature are
known:

– first buffer solution

pH 7,0 ± 0,5

– second buffer solution with

any pH value

* Rinse the sensor.
* Immerse the sensor in the first
buffer solution pH 7.0 ± 0.5.
* Wait for a stable measured
value.

Follow the instructions on the
display. As soon as a stable
measured value is reached, the
next display appears.

* Enter the pH value of the first
buffer solution.

Select the pH value of the first
buffer solution according to the
displayed temperature with

d

and confirm with

g

.

* Rinse the sensor.
* Immerse the sensor in the sec-
ond buffer solution.
* Wait for a stable measured
value.

Follow the instructions on the
display. As soon as a stable
measured value is reached, the
next display appears.

* Enter the pH value of the sec-
ond buffer solution.

Select the pH value of the second
buffer solution according to the
displayed temperature with

d

and confirm with

g

.

Successfully calibrated.
End of the CAL CON 2P
calibration.

The values determined for Slope
and Asymmetry potential are
displayed.
The calibration is complete.
Confirm with

g

.

The display returns to the
measured value display.