4 calibration with cal con 1p, Calibration with cal con 1p – YSI IQ SensorNet SensoLyt Sensor 700 IQ (SW) User Manual

Page 23

Advertising
background image

SensoLyt

®

700 IQ (SW)

Measuring / Operation

4 - 5

ba76006e01

01/2012

4.2.4

Calibration with CAL CON 1P

Messages displayed during the single-point calibration with
CAL CON 1P

Display

Explanation

* Have any buffer solution ready.

You can use any buffer solution
the pH value of which is known at
the current temperature.

* Rinse the sensor.
* Immerse the sensor in the buffer
solution.
* Wait for a stable measured
value.

Follow the instructions on the
display. As soon as a stable
measured value is reached, the
next display appears.

* Enter the pH value of the buffer
solution.

Enter the pH value with

d

.

Confirm entered data and
continue calibration

Confirm with

g

. The

measurement of the buffer
solution begins.

Successfully calibrated.
End of the CAL CON 1P
calibration.

The values determined for Slope
and Asymmetry potential are
displayed.
The calibration is complete.
Confirm with

g

.

The display returns to the
measured value display.