YSI IQ SensorNet SensoLyt Sensor 700 IQ (SW) User Manual
Page 34
Advertising
What to do if...
SensoLyt
®
700 IQ (SW)
7 - 2
ba76006e01
01/2012
Measurement provides
implausible measured
values
Cause
Remedy
– No calibration performed
– Calibrate
– Electrode not connected or
defective
– Check electrode and electrode
connection
– Electrode contaminated
– Clean electrode
– Liquid has penetrated the
sensor
– Sensor defective, send it back
– Instrument setting incorrect
– Correct instrument setting
(Measuring mode pH or mV)