4.2.7
Reactivation of the last valid calibration
The SensoLyt
®
700 IQ (SW) enables you to reactivate the last valid
calibration when needed. Thereby, it is possible to continue with
measurements, when a calibration failed or it is assumed that optimal
calibration conditions were not met.
Note
The reactivation of former calibration data is only a temporary solution.
Please consider, that thereby possibly wrong measurement values are
produced. Please check correct functioning of the sensor by performing
a function check and/ or a user calibration.
Reactivating the last
valid calibration data
1
Open the the setting table (see section 3.4).
2
Choose in the menue item Calibration the setting letzte gültige
and quit the setting table afterwards with Save and quit.