SensoLyt

®

700 IQ (SW)

Technical data

8 - 1

ba76006e01

01/2012

8

Technical data

8.1

Measurement characteristics

Measuring principle

Potentiometric measurement using a combination electrode;
Integrated microprocessor electronics, shielded 2-wire connection for
power and data transmission.

Measuring range

Resolution

Temperature

measurement

Temperature

compensation

8.2

Application characteristics

Temperature range

Allowed pH range of the

test sample

4 ... 12

pH

0.00 ... 14.00 pH (depending on the electrode)

ORP

-2000 mV ... +2000 mV
(depending on the electrode)

pH

0.01 pH

ORP

1 mV

Temperature sensor

integrated NTC

Measuring range

- 5 °C ... + 60 °C (23 ... 140 °F)

Accuracy

± 0.5 K

Resolution

0.1 K

Response time t

99

of the

temperature sensor

< 15 s

in the range 0 °C ... 60 °C (32 ... 140 °F)

Measuring medium

0 °C ... + 60 °C (32 ... 140 °F)

Storage/transport

- 5 °C ... + 65 °C (23 ... 149 °F)