YSI IQ SensorNet SensoLyt Sensor 700 IQ (SW) User Manual

Page 20

Advertising
background image

Measuring / Operation

SensoLyt

®

700 IQ (SW)

4 - 2

ba76006e01

01/2012

Maintenance condition

During calibration the sensor is in the so-called maintenance condition.
This means all linked outputs retain their momentary state. After
finishing calibration the maintenance condition has to be switched off
manually. For more detailed information on the maintenance condition
please refer to the respective IQ S

ENSOR

N

ET

system operating

manual.

General course of a

calibration on the

IQ S

ENSOR

N

ET

Generally, calibration on the IQ S

ENSOR

N

ET

is carried out as follows.

System specific details are given in the respective IQ S

ENSOR

N

ET

system operating manual.

Note
Before starting make sure the correct calibration procedure is set (see
section 3.4 S

ENSO

L

YT

® 700 IQ (SW)

SETTING

TABLE

).

1

Switch to the measured value display with

m

and select the

sensor to be calibrated.

2

Call up calibration with

c

.

The next step switches on the maintenance condition for the
sensor. A corresponding note appears on the display.

3

Confirm the note with

g

.

The maintenance condition is active.
The menu-guided calibration routine starts.
Follow the instructions on the display.
After the calibration routine is finished, the measured value
display appears again (the measured value flashes because
the sensor is still in the maintenance condition).

4

If the calibration was successful, bring the sensor into the
measuring position.

5

Wait for a stable measured value.

6

Switch off the maintenance condition.