YSI IQ SensorNet SensoLyt Sensor 700 IQ (SW) User Manual

Page 17

Advertising
background image

SensoLyt

®

700 IQ (SW)

Commissioning

3 - 7

ba76006e01

01/2012

Carrying out settings

Using

s

, switch from the measured value display to the main menu of

the settings. Then navigate to the setting menu (setting table) of the
sensor. The exact procedure is given in the relevant IQ S

ENSOR

N

ET

system operating manual.

active indicates that the sensor is being
calibrated. If Cancel is selected, the active
calibration procedure is canceled as soon
as the setting table is exited with Save and
quit
.

ORP shift
(only in mV
measuring mode)

-100 mV ... +100 mV

You can set the ORP zero point here.

Temperature
adjustment

-1.5 K ... +1.5 K

The temperature compensation function
enables the temperature sensor to be
balanced against a reference temperature
measurement (displacement of the zero
point by ±1.5 K).
Notes:
 Due to the thermal capacity of the

sensor, it is necessary to place it in a
container with at least 2 liters of water.

 Leave the sensor in this container for at

least 15 minutes, or in the case of
temperature differences between the
water and sensor > 10 K for at least 1
hour, while stirring occasionally. Then
carry out the balancing procedure.

Save and quit

The system confirms the saving of the
settings and the display switches to the
next higher level.

Quit

The display switches to the next higher
level without saving the new settings.