Measuring / Operation

SensoLyt

®

700 IQ (SW)

4 - 6

ba76006e01

01/2012

4.2.5

Calibration result

Calibration evaluation

After calibrating the system automatically evaluates the calibration data
and current state of the sensor. The asymmetry and slope are
evaluated separately. The values must be within the following ranges:

Slope:

-50 ... -62 mV/pH

Asymmetry:

-45 mV ... +45 mV

If one of both values is outside the specified range, the calibration is
evaluated as unsuccessful, i. e. the sensor could not be calibrated.

A calibration can have the following results:

Possible results of the

calibration

Note
Information on the contents and structure of the log book, and how to
call it up, is given in the L

OG

BOOK

chapter of the IQ S

ENSOR

N

ET

system operating manual.

Display after the
calibration

Log book entries
(meaning/actions)

Measured value display

Sensor was successfully calibrated.
For the calibration data, see the calibration
history.

"----"

Sensor could not be calibrated.
Sensor blocked for measurement.
– Service the sensor immediately

(see operating manual).

– View the calibration history.

– Check the calibration conditions and

calibration standard.