Measuring / Operation

SensoLyt

®

700 IQ (SW)

4 - 8

ba76006e01

01/2012

4.2.7

Reactivation of the last valid calibration

The SensoLyt

®

700 IQ (SW) enables you to reactivate the last valid

calibration when needed. Thereby, it is possible to continue with
measurements, when a calibration failed or it is assumed that optimal
calibration conditions were not met.

Note
The reactivation of former calibration data is only a temporary solution.
Please consider, that thereby possibly wrong measurement values are
produced. Please check correct functioning of the sensor by performing
a function check and/ or a user calibration.

Reactivating the last

valid calibration data

1

Open the the setting table (see section 3.4).

2

Choose in the menue item Calibration the setting letzte gültige
and quit the setting table afterwards with Save and quit.