10 indexes, 1 explanation of the messages, 1 error messages – YSI IQ SensorNet SensoLyt Sensor 700 IQ (SW) User Manual

Page 41: 10 indexes -1, 1 explanation of the messages -1, 1 error messages -1

Advertising
background image

SensoLyt

®

700 IQ (SW)

Indexes

10 - 1

ba76006e01

01/2012

10

Indexes

10.1

Explanation of the messages

This chapter contains a list of all the message codes and related
message texts that can occur in the log book of the IQ S

ENSOR

N

ET

system for the SensoLyt

®

700 IQ (SW) sensor.

Note
Information on the contents and structure of the log book, and how to
call it up, is given in the L

OG

BOOK

chapter of the IQ S

ENSOR

N

ET

system operating manual.

Note
All Message codes of the SensoLyt

®

700 IQ (SW) end up with "311".

10.1.1 Error messages

Message code

Message text

EA1311

Meas. range exceeded or undercut
* Check process
* Select other meas. range

EA2311

Sensor temperature too high!
* Check process and application

EA3311

Sensor temperature too low!
* Check process and application

EC1311

Sensor could not be calibrated,
Sensor blocked for measurement
* Check calibration conditions and calibration standard
* View calibration history
* Service sensor immediately
(see operating manual)

EI1311

Operational voltage too low
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components, replace components

EI2311

Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components,
replace components