Sensor storage, Troubleshooting – YSI ProOBOD User Manual

Page 10

Advertising
background image

12

13

sensOr stOrage
SHORT-TERM STORAGE

When the ProOBOD is not in use, it MUST BE STORED IN A MOIST

ENVIRONMENT, i.e., the sensor either immersed in water or in water-saturated

air. If the sensor cap is allowed to dry out by exposure to dry air, it is likely

to drift slightly at the beginning of its next use unless it is rehydrated. If this

occurs, follow the rehydrating instructions in this manual.

For short-term storage (<30 days), place the probe in a BOD bottle that contains

a small amount of clean water (approximately 40 ml). This will provide a 100%

water-saturated air environment.

LONG-TERM STORAGE

For long-term storage (>30 days), remove the batteries from the instrument.

Moisten the sponge in the clear protective plastic cap that was originally provided

with the probe with clean water. Place the clear cap over the sensor with the

sensor cap installed. Inspect the sponge every 30 days to make sure it is still

moist. If you no longer have the clear protective cap, place the probe in a BOD

bottle that contains a small amount of clean water (approximately 40 ml).

Alternatively, you can place the sensor with sensor cap installed directly in water

in a BOD bottle, making sure that the water does not evaporate over time.

Recommended Long-term Storage ambient temperature: -5 to 50°C (23 to 122°F)

T R O U B L E S H O O T I N G

DissOlveD Oxygen reaDings

Erroneous dissolved oxygen readings typically indicate a need to clean the sensor

cap, replace the sensor cap, and/or recalibrate the instrument. First, verify the

sensor is properly setup in the Probe menu. Next, clean the sensor cap following

the instructions in the Maintenance and Storage section of this manual and

then perform a calibration. If erroneous readings persist, follow the steps on

the following page to inspect the sensor cap for damage and then attempt to

recalibrate the instrument. If the problem continues, try to rehydrate the sensor

cap and then recalibrate. If you are still getting erroneous dissolved oxygen

readings, try replacing the sensor cap and then recalibrate. If the erroneous

readings continue, contact YSI Technical Support to help determine the next

step.

INSPECTING THE SENSOR CAP FOR DAMAGE

Caution: Avoid touching the sensing end of the sensor

cap during the following maintenance procedures.

If DO measurements seem to be in error or are jumpy, remove the sensor cap

from the sensor by grasping the probe body with one hand and then rotating the

sensor cap counterclockwise until it is completely free. Do not use any tools for

the removal of the cap.

Inspect the sensor cap assembly for any cracks or damage. If damage has

occurred, contact YSI Customer Service to order a replacement sensor cap (YSI

item number 626458).

Inspect the o-ring on the probe for damage. If there is any indication of damage,

carefully remove the o-ring and contact YSI Technical Support to obtain a new

o-ring and/or gasket. Avoid using tools to remove the o-ring as damage to the

sealing surfaces could result.

Before reinstalling the sensor cap, make sure that the cavity is completely clean

and dry before proceeding with the installation. If water is found, dry the cavity

with lens tissue. Finally, clean and dry the clear optical DO sensor window on

the end of the sensor with a lens tissue.

After reinstalling the sensor cap, perform a calibration and then reevaluate

the quality of the dissolved oxygen measurements. If problems persist, try

rehydrating or replacing the sensor cap.

help

While using of the ProODO instrument, press the Question

key from any

screen to view helpful messages directly on the display.

errOr/statUs messages

If readings for temperature or dissolved oxygen are over or under range, you

will see a series of +++++ or ----- respectively and an error message along the

bottom of the display. A series of ????? indicates that a parameter, for example

DO mg/L, cannot be calculated. This could be caused by a connection issue

between the cable and instrument. The following are some of the potential error

messages:

u

Advertising