Getting started, Unpacking the obod probe – YSI ProOBOD User Manual

Page 5

Advertising
background image

2

3

G E T T I N G S T A R T E D

Throughout the manual, the term “probe” refers to the end of the cable where

the sensor, stir motor, and stirrer are located. The term “sensor” refers to the

Optical Dissolved Oxygen sensing portion of the probe assembly and the term

“sensor cap” refers to the removable sensing cap that is replaced about once per

year (Figure 1).

Figure 1.

Unpacking the OBOD prOBe

Each ProOBOD probe and replacement OBOD sensor cap

includes an instruction sheet with important information

unique and specific to each individual sensing cap.

Carefully unpack the instrument and accessories and inspect for damage.

Compare received parts with items on the packing list. If any parts or materials

are damaged or missing, contact YSI Customer Service at 800-897-4151 (+1 937

767-7241) or the authorized YSI distributor from whom the instrument was

purchased.

OBOD Sensor Cap,

shown here removed

from probe

Probe

Stir paddle

Optical DO sensor

Stirrer toggle switch,

turns stir paddle

on and off

Temperature sensor

i

Remove the ProOBOD cable/probe assembly from the shipping container and

locate the instruction sheet that is included. This instruction sheet is important

because it contains the calibration coefficients for your sensor cap. After using

this sheet for general probe setup, be sure to store it in a safe place in case you

need to reload these calibration coefficients in the unlikely event that they are

ever deleted from the instrument and probe.

Note - A new cable/probe assembly already has a

sensor cap installed and the sensor cap coefficients

are preloaded into the probe at the factory.

Preparing the probe for the first time:

A new ProOBOD probe will have a clear protective cover placed over the

1.

sensor cap to ensure the cap remains moist during shipment. Carefully

remove this clear cover from the end of the probe by pulling it straight off

the sensor. Save the clear cover along with the sponge in case you want to

use it later for long term storage of the probe.

Place the probe in a standard BOD bottle that contains a small amount of

2.

clean water (approximately 40 ml). The sensor should not be immersed in

water but rather in the air above the water. The purpose is to create a 100%

water-saturated air environment for the sensor which is ideal for calibration

and storage when not in use.

It is important to always keep your sensor in a moist

environment so the sensor cap does not dry out. (See

Maintenance and Storage for more information.)

u

i

Advertising