Proobod operation – YSI ProOBOD User Manual

Page 6

Advertising
background image

4

5

P r o O B O D O P E R A T I O N

cOnnecting prOBe/caBle assemBly tO instrUment

To connect the cable, align the keys on the cable connector to the slots on the

instrument connector. Push together firmly, then twist the outer ring until it

locks into place (Figure 2). This MS-8 (Military Spec) connection is water-

proof.

Figure 2. Note the keyed connector. The cable and instrument connectors

can only be mated once the keyed sections are properly aligned.

cOnfigUring the p

ro

ODO instrUment

On the ProODO instrument, press the Probe

key. Highlight DO and press

enter. Enabled allows you to enable or disable the DO sensor. Highlight Enabled

and press enter to activate or deactivate the dissolved oxygen sensor.

The DO Setup menu also displays the OBOD probe’s SW Version and Serial #.

This information is programmed into the probe at the factory and digitally sent

to the instrument when the cable is connected.

Within the DO Probe Setup menu, the Sensor Cap submenu allows you

to view and enter information specific to the sensor cap installed on the

probe. Highlight Sensor Cap and press enter to view the sensor cap’s Serial #,

Temperature Coefficient, and Sensor Cap Coefficients. This information is also

programmed into the probe at the factory and sent to the instrument when the

cable is connected.

The Sensor Cap Coefficients need to be updated when the sensor cap is replaced.

The frequency of sensor cap replacement is dependent on use. For example,

when the probe is powered on for approximately 4 hours per day, 5 days a week,

the sensor cap will need replaced about once per year. If the probe is powered

on more than this, the sensor cap will need replaced more often. To extend the

sensor cap’s usable life, turn the instrument off when not in use. For example,

turn the instrument off over night, but not between readings. The sensor cap

will also need replaced if it is damaged or cracked. See the Troubleshooting

section of this manual for instructions on how to determine if the sensor cap

needs replaced. See the Sensor Maintenance - Dissolved Oxygen section of

this manual or the instruction sheet included with the replacement sensor cap

for instructions on updating the Sensor Cap Coefficients. When updating the

Coefficients, the sensor cap serial # will be updated automatically based on your

entries.

The Temperature Coefficient should not be modified unless instructed by YSI

Technical Support.

For additional information about the ProODO’s menus and operation, please

refer to the ProODO User Manual.

pOwering the stirrer

To power the stirrer, locate the power supply that was included with the

ProOBOD. The power supply for the USA/Japan probe is labeled model number

6260. The power supply for the international probe is labeled model number

6261. Next, connect the power supply to the probe’s power connector and then

plug the power supply into an outlet. The international ProOBOD is shipped

with outlet adapters to connect the power supply to a variety of outlets. The

stirrer can be turned on and off with the toggle switch located at the top of the

probe (Figure 1).

caliBratiOn - DissOlveD Oxygen

The ProOBOD sensor is an optical luminescent sensor which has greater stability

and is less susceptible to calibration drift than traditional electrochemical

sensors. This increased stability means that the instrument may hold its

calibration for many months. However, for the highest data accuracy, YSI

recommends verifying the instrument’s calibration on a daily basis. To verify

the instrument’s calibration, place the sensor in its calibration environment and

check to see that the DO% is reading its calibration value based on the current

barometric pressure. Refer to Appendix A for a list of DO% calibration values

and corresponding barometric pressure readings.

The ProODO instrument with ProOBOD probe offers several options for

performing a dissolved oxygen calibration. For both ease of use and accuracy,

YSI recommends performing a DO % water-saturated air calibration as described

below. For information on other acceptable calibration techniques, including

how to perform a 2-point calibration with a zero solution, please refer to the

ProODO User Manual.

Advertising