Temperature taking measurements – YSI ProOBOD User Manual

Page 7

Advertising
background image

6

7

It is not necessary to calibrate in both % and mg/L or ppm.

Calibrating in % will simultaneously calibrate

mg/L and ppm and vice versa.

CALIBRATING DO% IN WATER-SATURATED AIR:

1-POinT CalibraTiOn

Make sure there are no water droplets on the OBOD sensor cap or temperature

sensor and then place the probe in a BOD bottle that contains a small amount

of clean water (approximately 40 ml). The BOD bottle should be clean since

bacterial growth may consume or produce oxygen which would interfere with

the calibration. Make sure the DO and temperature sensors are not immersed in

water. The purpose is to create a 100% water-saturated air environment. Next,

wait approximately 5 to 10 minutes or until the DO and temperature readings

are stable. This will allow for the air in the BOD bottle to become completely

water-saturated. The design of the probe ensures venting to the atmosphere

when placed in a typical BOD bottle which is required to create the 100% water-

saturated air calibration environment.

Press the Calibration

key on the ProODO instrument. Highlight DO and

press enter. Highlight DO % and press enter to confirm.

The instrument will use the reading from the

internal barometer during calibration and will

display this value in brackets at the top of the

display. Highlight the barometer value and

press enter to adjust it if needed. Note - the

barometer should be reading “true” barometric

pressure. If the barometer reading is incorrect,

perform a barometer calibration. See the

Barometer section in the ProODO User Manual

for more information on “true” barometric

pressure and on how to perform a barometer

calibration. If the barometer reading is acceptable, there is no need to change it

or perform a barometer calibration.

Wait for the temperature and DO% values under “Actual Readings” to stabilize,

then highlight accept Calibration and press enter to calibrate. If User Field 1
or 2 are enabled, you will be prompted to select the fields and then press Cal

to complete the calibration. The message line at the bottom of the screen will

display “Calibrating Channel...” followed by “Calibration Successful”. Press Esc

to cancel the calibration.

i

temperatUre

All ProOBOD probes have built-in temperature sensors. Temperature calibration
is not required nor is it available. To set the units, press the Probe

key on

the ProODO instrument and select Display. Highlight Temperature and press

enter. Highlight the desired temperature units of °F, °C, or K and press enter to

confirm the selection.

taking measUrements

Important: The ProOBOD is designed for laboratory use, not

field use. It is not waterproof and should never be immersed

in a sample past the taper on the stem of the probe

To take readings, insert the probe into the BOD sample bottle. The DO and

1.

temperature sensors should be immersed in the sample.

Turn the probe stirrer on with the toggle switch located on the top of the

2.

probe (Figure 1). The stirrer helps keep solids from settling at the bottom

of the sample bottle and quickens the sensor’s response.

Allow the temperature readings to stabilize and wait approximately 20-25

3.

seconds for the DO readings to stabilize.

Log One Sample will be highlighted on the ProODO instrument when it is

4.

in Run mode. Press Enter to open a logging submenu. Highlight Sites or

Folders and press Enter to select a site or folder to log the sample to.

Once the Site and/or Folder name is selected, highlight

5.

log now and press

Enter. The instrument will confirm that the data point was successfully

logged.

If you would like to log at a specific interval vs. logging one sample at a time

6.

or vice versa, press the System

key, then highlight logging and press

Enter. Select Continuous Mode and adjust the time interval if necessary.

On the Run screen, the option to log will change from Log One Sample to

Start Logging based on the time interval entered in the Logging Menu.

Note – When utilizing the Manual sampling mode, the

Continuous Logging Interval must be set to 10 seconds or

greater. For additional information on logging data and the

sampling mode options, refer to the ProODO User manual.

u

i

Advertising