Maintenance and storage – YSI ProOBOD User Manual

Page 8

Advertising
background image

8

9

M A I N T E N A N C E A N D S T O R A G E

This section describes the proper procedures for care, maintenance and storage

of the ProOBOD probe.

Important: Do not attempt to access the probe motor

assembly or open the probe body. Doing so will void

any remaining warranty. The probe body should only

be opened by a YSI Authorized Service Center.

UpDating instrUment anD prOBe firmware

The instrument and probe firmware can be updated via www.ysi.com/software.

There you will find the new firmware files and instructions on how to update the

instrument and/or probe.

sensOr maintenance - DissOlveD Oxygen
CLEANING THE OBOD SENSOR CAP

The Sensor Cap should be kept clean since some types of fouling may consume

or produce oxygen which could affect the dissolved oxygen measurements. To

clean the Sensor Cap, gently wipe away any fouling with a lens cleaning tissue

that has been moistened with water. Do not use a coarse towel or cloth or organic

solvents to clean the Sensor Cap. Using a coarse towel or organic solvent to clean

the Sensor Cap may cause permanent damage to the cap. For example, alcohol

will dissolve the outer paint layer and other organic solvents will likely dissolve

the dye in the cap. If the sensor cap is damaged, it must be replaced.

OBOD SENSOR CAP REPLACEMENT

The frequency of sensor cap replacement is dependent on use. For example,

when the probe is powered on for approximately 4 hours per day, 5 days a week,

the sensor cap will need replaced about once per year. If the probe is powered

on more than this, the sensor cap will need replaced more often. To extend the

sensor cap’s usable life, turn the instrument off when not in use. For example,

turn the instrument off over night, but not between readings. The sensor cap

will also need replaced if it is damaged or cracked. See the Troubleshooting

section of this manual for instructions on how to determine if the sensor cap

needs replaced.

When replacing a Sensor Cap, the Sensor Cap Coefficients must be manually

updated in the ProODO instrument. The instruction sheet shipped with the

replacement OBOD sensor cap includes the calibration coefficients that are

specific to your Sensor Cap.

u

IMPORTANT - Be sure to save the OBOD Sensor Cap instruction

sheet in case you need to reload the calibration coefficients.

The replacement OBOD Sensor Cap is shipped in a humidified container and

the package should not be opened until the cap is needed for use. Once the

sensor cap has been installed on the OBOD probe as described in this section, it

is important to keep the sensor in a 100% humid environment. Therefore, the

ProOBOD sensor should be stored in either a BOD bottle that contains a small

amount of clean water (approximately 40 ml) or in a BOD bottle that is filled

with clean water so the sensor cap is immersed in water, see

Sensor Storage for

more information. If the sensor dries out, refer to the Rehydration procedure in

this manual for instructions on how to recondition the sensor cap.

Refer to Figure 3 below when following the instructions for replacing the cap.

Figure 3.

Caution: Avoid touching the sensing end of the sensor

cap during the following maintenance procedures.

Remove the stir paddle from the probe by pulling it straight out.

1.

Remove the old sensor cap from the probe by grasping the probe body

2.

with one hand and then rotating the sensor cap counterclockwise until it is

completely free. Do not use any tools for this procedure.

Carefully remove the o-ring and discard it. Do not use any tools to remove

3.

the o-ring.

Clean the o-ring seat on the probe by first wiping off old o-ring lubricant

4.

with a dry lens cleaning tissue. Next, clean away any build-up with a water-

moistened lens tissue and then dry with another lens tissue.

Locate the o-ring supplied in the Sensor Cap replacement kit and install it

5.

on the probe. Do not use any tools to replace the o-ring. Be careful not to

touch the clear optical DO sensor window. Any fingerprints on the optical

u

O-ring

OBOD Sensor

Cap, shown

here removed

from probe

u

Advertising