Restore default calibration values, Probe specifications – YSI ProOBOD User Manual

Page 11

Advertising
background image

14

15

Message

Description and Recommended Action

Probe Temp over range

Temperature is over

50°C or reading

erroneously. Check cable connection

and ensure temperature sensor is clean.

Probe Temp under range

Temperature is under

-5°C or reading

erroneously. Check cable connection

and ensure temperature sensor is clean.

DO over range

DO% saturation is over 550%. Check sensor

cap and recalibrate.

DO under range

DO% saturation is under -5%. Check sensor

cap and recalibrate.

Barometer over range

Barometric pressure is reading over 988

mmHg. Calibrate barometer.

Barometer under range

Barometric pressure is reading under 375

mmHg. Calibrate barometer.

ODO Communications

Error

No communication between the instrument

and cable. Check cable connection.

Clock Battery Low

Internal battery for real time clock has low

voltage. Contact Technical Support.

Measurements Locked!

Measurements are held in Manual Sampling

mode. Select Update Measurements or Log

Held Data.

illegal Value may appear during alpha/numeric entry on the message line. This

only appears if the values entered do not match the formatting. This will also

appear in the GLP security area if the password is incorrect.

restOre DefaUlt caliBratiOn valUes

Occasionally, the instrument may need to have the factory calibration default

values restored. To perform a factory reset, press Calibrate

, highlight

restore Default Cal and press enter. Select the parameter you wish to restore,

either DO or Barometer, and press enter. After selecting barometer or DO, you

will be asked to confirm the operation. Highlight Yes and press enter.

P R O B E S P E C I F I C A T I O N S

The specifications listed below are system specifications for the ProOBOD probe

when used with a ProODO instrument. These specifications are subject to

change without notice. For the most recent specifications and for specifications

on the ProODO instrument, please visit www.ysi.com.

Probe (ProObOD probe and cable assembly)
Temperature

Operating Range

10 - 40°C

Water Resistance

IP-65

Cable Length

2 meters (6.56 feet)

Dissolved Oxygen
Sensor Type

Optical, Life-time Luminescent sensor

Range

0-50 mg/L

0-500 % Saturation

Accuracy (mg/L)

0 to 20 mg/L: ± 0.1 mg/L or ± 1% of

reading, whichever is greater.

20 to 50 mg/L: ± 15% of the reading

Accuracy

(% Saturation)

0 to 200% air saturation: ± 1% of the reading or ±

1% air saturation, whichever is greater.

200 to 500% air saturation: ± 15% of the reading

Resolution (mg/L) 0.01 or 0.1 mg/L (auto-scaling)
Resolution (%

Saturation)

0.1% air saturation

Response Time

T95 = 20-25 seconds with stirring;

T95 = 40 seconds without stirring.

Temperature
Range

-5 to 50°C

Accuracy

± 0.2°C

Resolution

0.1°C