3 sensor-independent settings, 1 system, Sensor-independent settings – YSI TruLab 1310P User Manual

Page 43: System, Section 7.3.1 s, Ystem

Advertising
background image

TruLab pH 1310

Settings

ba76156e01

03/2014

43

ter in the measured value display and press the <F1>/[menu] or <MENU/
ENTER>
key. After completing the settings, switch to the measured value dis-
play with <M>.
Default settings are printed in bold.

7.3

Sensor-independent settings

7.3.1 System

To open the Storage & config menu, press the <F1__>[Menu] key in the mea-
sured value display. After completing the settings, switch to the measured
value display with <M>.
Default settings are printed in bold.

Menu item

Possible
setting

Description

Man. temperature

-25 ... +25
... +130 °C

Entry of the manually determined tem-
perature. For measurements without
temperature sensor only.

Temperature unit

°C
°F

Temperature unit,

degrees Celsius or degrees Fahrenheit.
All temperatures are displayed with the
selected unit.

Resolution mV

0.1
1

Resolution of the mV display

Stability control

On
Off

Switches on or off the automatic stability
control during measurement (see sec-
tion Stability control (AutoRead) &
HOLD function, page 22
)

Reset

-

Resets all sensor settings to the delivery
condition (see section 7.4.1 R

ESETTING

THE

MEASUREMENT

SETTINGS

, page 45).

Menu item

Possible set-
ting

Description

System / General /
Language

Deutsch
English
(further)

Selects the menu language

System / General /
Beep

On
Off

Switches on/off the beep on
keystroke

System / General /
Illumination

Auto
On
Off

Switches the display illumina-
tion on/off

System / General /
Contrast

0 ... 50 ... 100 %

Changing the display contrast

Advertising
This manual is related to the following products: