TruLab pH 1310
Settings
ba76156e01
03/2014
43
ter in the measured value display and press the <F1>/[menu] or <MENU/
ENTER> key. After completing the settings, switch to the measured value dis-
play with <M>.
Default settings are printed in bold.
7.3
Sensor-independent settings
7.3.1 System
To open the Storage & config menu, press the <F1__>[Menu] key in the mea-
sured value display. After completing the settings, switch to the measured
value display with <M>.
Default settings are printed in bold.
Menu item
Possible
setting
Description
Man. temperature
-25 ... +25
... +130 °C
Entry of the manually determined tem-
perature. For measurements without
temperature sensor only.
Temperature unit
°C
°F
Temperature unit,
degrees Celsius or degrees Fahrenheit.
All temperatures are displayed with the
selected unit.
Resolution mV
0.1
1
Resolution of the mV display
Stability control
On
Off
Switches on or off the automatic stability
control during measurement (see sec-
tion Stability control (AutoRead) &
HOLD function, page 22 )
Reset
-
Resets all sensor settings to the delivery
condition (see section 7.4.1 R
Menu item
Possible set-
ting
Description
System / General /
Language
Deutsch
English
(further)
Selects the menu language
System / General /
Beep
On
Off
Switches on/off the beep on
keystroke
System / General /
Illumination
Auto
On
Off
Switches the display illumina-
tion on/off
System / General /
Contrast
0 ... 50 ... 100 %
Changing the display contrast