Operation

Turb 430 IR/T

36

ba75507e03

07/2006

4.5.4

Settings for turbidity measurements

Overview

For turbidity measurements, the following settings are possible in the
menu, Configuration / Turbidity:

! Calibration record (display, print)
! Entering the Calibration interval
! Reset

Settings/functions

The settings are given in the menu, Configuration / Turbidity.
Move to the Configuration menu with the <MENU> key.

4.5.5

Calibration

When to calibrate?

! After the calibration interval has expired
! With a temperature change

Calibration procedures

and calibration

standards

For the menu-guided three-point calibration you need the following
three calibration standards in the mentioned order:

Menu item

Possible
setting

Description

Calibration record

-

Displays the calibration
record of the last
calibration.

Calibration interval

1 ... 999 d

Calibration interval for
turbidity measurement (in
days).

If the calibration interval
has expired, the meter
reminds you to calibrate
before each measurement.

Reset

Reset all settings for the
Turbidity measuring mode
(see section 4.8.2)

Standard no.

FNU/NTU

1

1000

2

10,0

3

0,02