If the memory is full

You can erase the entire memory (see section 4.6.5), or overwrite the
oldest dataset with the next storing procedure.
A security prompt appears before a dataset is overwritten.

4.6.2

Filtering measurement datsets

The functions to display and download stored measurement datsets
(see section 4.4.1) refer to all stored measurement datsets that
correspond to the specified filter criteria.

The settings are given in the menu, Configuration / System / Measured
value memory
/ Data filter.
Move to the Configuration menu with the <MENU> key.

Data filter

Menu item

Setting/function

Description

Filter

Filter criteria:

No filter

Data filter switched off

ID

Selection according to ID
number

Date

Selection according to
period

ID + Date

Selection according to
period and ID number

ID

Entry of filter criteria

These menu items are
made visible by selecting
the filter criteria in the Filter
menu.

Date