Xylem TURB 430 PORTABLE TURBIDITY METER User Manual

Page 38

Advertising
background image

Operation

Turb 430 IR/T

38

ba75507e03

07/2006

Note
Before measuring the third calibration standard of 0.02 FNU/NTU you
can exit the calibration with <ESC> at any time.
The new calibration data is discarded. The old calibration data is further
used.

3

Align the cell:
! Marked cell:

– Align the marking on the cell cap with the marking on the

cell shaft.

– Press and hold the <START/ENTER> key until the

measured value is displayed.

! Unmarked cell (see page 33)

– Press and hold the <START/ENTER> key.

– Slowly and in small steps turn the cell by one complete

rotation (by 360 °).

– After each step wait for a short time until the displayed

measured value is stable.

– Turn the cell back to the position with the lowest

measured value.

4

Release the <START/ENTER> key.

Measurement of the calibration standard begins.

5

Repeat the steps 4 - 6 with the calibration standards, 10.0 FNU/
NTU and 0.02 FNU/NTU.

After measuring the 0.02 FNU/NTU calibration standard, the
calibration result is displayed.
Calibration is completed.

6

Confirm the calibration result with <START/ENTER>.

The calibration record is displayed.

7

Confirm the calibration record with <START/ENTER>.

The display shows instructions for the first measurement.

i

Turb.

= 1000 FNU/NTU

i Start calibration
by releasing <START>

Turbidity \ Calibration